[导读]设备名称:GE MR机器型号:EXCITE HD 1.5T关 键 词:错误代码2223621、2223849 系统启动超时故障描述(信息查看):1.机器在扫描期间报错:224862The GP is reportin
设备名称:GE MR
机器型号:EXCITE HD 1.5T
关 键 词:错误代码2223621、2223849 系统启动超时
故障描述(信息查看):
1.机器在扫描期间报错:224862
The GP is reporting a SGA Power Supply Under Voltage Fault.
2.之后做TPS不通,重启
3.系统超时,依次报错:
2223621
Internal Software ERROR.Can’t callback on undefined NMRID:SCT
2218574
Communications have failed. Please try again.
2252106
Diagnostic software error.
Failed to determine hardware state for AGP board within 2- seconds.
The board may be busy loading vxWorks or may be hung.
Also check the serial cable for the board and the serial terminal server.
2223849
AGP/SCP/APS EPROM poewr-up boot failed.Verify that MCP750 boards are
installed and seated properly for APS and AGP.
故障诊断(分析):
1.从主机上ping APS板通,ping AGP板不通,ping SCP板通,ping Term
Server通;
2.去设备间发现AGP板上的四个灯都不亮,正常情况下AGP板和APS板上的灯应该同闪烁(如下图);
1.JPEG
3. 随后用笔记本连接Network Switch,并pingAPS板,AGP板,SCP板,Term Server只有AGP板不通;
4. 对Network Switch上与APS,AGP,SCP相连接的网线依次进行检查正常;
5. 随后对Term Server断电重启设置并测量电源电压12V均正常;
6. 最后对MGD柜进行最小化,并对各个板子逐一插入检查,只有当IRF板插入后,AGP板上的灯就会不亮且无论从主机还是Network
Switch上pingAGP板都不通。
故障解决:
更换IRF板后,观察AGP板和APS板上的灯同步,且都可以ping通,重启系统正常,TPS通过,扫描正常。
维修心得:
IRF板故障引起系统不能正常启机,TPS不通。